Używamy plików cookies, aby zapewnić najlepsze doświadczenia na naszej stronie. · Polityka prywatności
Tightening an implant screw "by feel" is a clinical error. Both undertightening and overtightening have consequences – loosening or fracture of the screw. Below, we've compiled parameters and procedures from daily clinical and laboratory practice.
Tightening an implant screw "by feel" is a clinical error. Both undertightening and overtightening have consequences – loosening or fracture of the screw. Below, we've compiled parameters and procedures from daily clinical and laboratory practice.
Torque is the rotational force expressed in Newton centimeters (Ncm) that must be applied to a screw to achieve stable preload without the risk of damage. Each implant system manufacturer provides recommended values in the technical data sheet.
The value is not the same for all screws – it depends on the system and the type of connection (external/internal, locking taper).
| System | Abutment Screw | Crown Screw | Connection Type |
|---|---|---|---|
| Straumann BL | 35 Ncm | 15 Ncm | Locking Taper |
| Nobel Biocare (NP/RP) | 35 Ncm | 15 Ncm | Conical |
| Osstem TSIII | 30 Ncm | 15 Ncm | Internal Hex |
| Biomet 3i | 20 Ncm | 20 Ncm | External Hex |
| Dentsply Ankylos | 15 Ncm | 15 Ncm | Locking Taper |
Always verify with the manufacturer's technical data sheet for the specific implant lot.
Undertightening – the screw does not achieve the required preload. Result: loosening under functional forces within the first few weeks of use.
Overtightening – exceeding the manufacturer's recommended value can cause permanent deformation of the thread or screw fracture. With locking taper connections, the risk is particularly high.
Lack of documentation – during subsequent disassembly (aesthetic revision, repair), you do not know the force with which the screw was seated. Lack of documentation carries the risk of component damage during removal.
Loose screw – what to do? Loosening within the first 6 months is the most common time for the problem to occur. Procedure: remove the prosthesis, replace the screw (do not re-tighten the loosened one), tighten to the correct torque, document. Loosening can also indicate occlusal overload – check occlusion before re-seating.
LABORATORY PERSPECTIVE
Contact deltalabs. — we will advise on the best solution for your case.
Get in touch — we'll discuss your case and find the optimal solution.
3SHAPE · ITERO · MEDIT · DENTSPLY SIRONA